l、移门限位报警,可时刻提醒操作者。
Sliding door stopper alarm can remind the operator.
2、排风有高效过滤器,保证排放的空气符合国家卫生要求。
A dedicated and efficient exhaust filters to ensure air emissions meet state health requirements.
3、配备防水插座和排污接口为操作者提供*方便。
Equipped with special water and sewage outlet operator interface provides great convenience
4、既保证人员又保证受试样本的安全。
To ensure the safety of personnel and tested samples
参数/型号 Parameter\Model | BSC-1300-Ⅱ-A | BSC-1300-Ⅱ-B1 | BSC-1300-Ⅱ-B2 |
洁净等级 Clean grade | ISO5级(class100级) 100 Grade@≥0.5μm (209E) | ||
菌落数 Number of bacteria | ≤0.5个/皿.时(¢90mm培养平皿) ≤0.5per utensil.hour(¢90mm utensil) | ||
气密度 Air tightness | 500Pa压力下泄漏量≤10%(30分钟内) 500Pa leakage rate≤10%(in 30min) | ||
噪 音noise | ≤62dB | ||
振动半峰值 Half peak value of shaking of shaking | ≤5μm | ||
照 度 Illumination | ≥500Lx | ||
电 源 Power supply | AC,Ф1220V/50Hz | ||
zui大功耗 Max. power consumption | 0.8KW | ||
送风风速 Equiblast Speed | 0.25~0.45m/s,可调 0.25~0.45m/s(adjustable) | ||
吸入风量 Equiblast air volume | ≥0.5m/s | ||
重 量Weight | 280kg | 300kg | 350kg |
工作区尺寸(W1×D1×H1) Dimension of working area(W1×D1×H1) | 1300*600*650mm | 1300*600*650mm | 1300*600*650mm |
外形尺寸(W1×D1×H1) Overall dimension(W1×D1×H1) | 1500*800*2230mm | 1500*800*2230mm | 1500*800*2230mm |
送风过滤器规格和数量 Specification and number of air supply filter | 1260*520*70*① | 1260*520*70*② | 1260*520*70*② |
排风过滤器规格和数量 Specification and number of air discharge filter | 610*305*90*① | 968*484*90*① | 1220*610*90*① |
荧光灯/紫外灯规格和数量Specification and number of light from firefly/ ultraviolet light | 40W*③/30W*① | 40W*③/30W*① | 40W*③/30W*① |
备 注 Remark | 30%排风 30% exhaust air rate | 70%排风 70% exhaust air rate | *排风 * exhaust air rate |